4. Prepare Standards

General Instructions

n

It is essential to prepare standards exactly as described in this section.

Incorrect preparation may lead to low signal or variable measurements
from plate to plate

n

The peel-off sticker provided with the standards lists the most

concentrated point on the standard curve (S1). Enter this information
into Bio-Plex Manager

software as instructed in section 8

Prepare a Diluent for Standards

1. Refer to Table 4 for recommended diluents based on different sample

types. As a general rule, reconstitute and dilute standards in a diluent
similar to the final sample type or sample matrix.

2. If samples are serum or plasma, mix 1 volume of Bio-Plex

®

standard diluent with 3 volumes of Bio-Plex sample diluent (each
supplied in the kit). The resulting solution is used for reconstitution
and subsequent dilution of standards. This results in a serum-matrix
based diluent that mimics the matrix in 1:16 diluted serum and
plasma samples.

Sample Type

Diluent for Standard

Add BSA

Serum and plasma

Standard/sample diluent mix (1:3)

None

Culture media, with serum

Culture media

None

Culture media, serum-free

Culture media

To 0.5% final* (w/v)

Lavage, lysate, other fluids

Sample diluent

To 0.5% final* (w/v)

* At least 0.5% final w/v BSA is recommended to stabilize analytes and reduce adsorption

to labware.

Table 4. Summary of recommended diluents for standards.