Bio-Rad Bio-Plex Pro™ TGF-β Assays User Manual

Page 35

Advertising
background image

33

Possible Causes

Low Bead Count

Miscalculation of bead dilution

Beads clumped in multiplex
bead stock tube

Vacuum on for too long when
aspirating buffer from wells

Filter plate not shaken enough
before incubation steps and prior
to reading

Reader is clogged

Low Signal or Poor Sensitivity

Standards reconstituted
incorrectly

Detection antibody or

streptavidin-PE diluted incorrectly

Possible Solutions

Check your calculations and be
careful to add the correct volumes.

Vortex for 30 sec at medium speed
before aliquoting beads.

Do not apply vacuum to the filter
plate for longer than 10 sec after the
buffer is completely drained from
each well.

Shake the filter plate at
850 ± 50 rpm for 30 sec before
incubation steps and immediately
before reading the plate.

Refer to the troubleshooting guide
in the Bio-Plex system hardware
instruction manual (bulletin
#10005042).

Follow the standard preparation
instructions carefully.

Check your calculations and be
careful to add the correct volumes.

Advertising