Rp1 (pmt) setting for standard curves, The bio-plex – Bio-Rad Bio-Plex Pro™ TGF-β Assays User Manual

Page 16

Advertising
background image

14

Fig. 3. Preparing a fourfold dilution series with a single reconstituted standard.

RP1 (PMT) Setting for Standard Curves

The Bio-Plex

®

200 and 3D systems have two RP1 (PMT or photomultiplier

tube) setting options, while the Bio-Plex

®

MAGPIX™ has no PMT

and therefore no PMT setting options. Instead, MAGPIX uses default
instrument settings similar to low PMT on the Bio-Plex 200 (Table 5).

The Bio-Plex Pro TGF-b assays were developed on the low PMT setting
using the Bio-Plex 200 system. Protocols using alternative standard
dilution series or PMT settings should be validated by the end user.

One

Reconstituted

Standard

72 150 150 150 150 150 150 150 150

S1 S2 S3 S4 S5 S6 S7 S8 Blank

Diluent, µl

128 50 50 50 50 50 50 50

Transfer Volume, µl

Instrument

RP1 (PMT)

Bio-Plex 100, 200*

Low, high

Bio-Plex 3D*

Standard, enhanced

Bio-Plex MAGPIX*

N/A, use default instrument settings

* Or similar Luminex-based system.

Table 5. Overview of PMT setting options on Bio-Plex systems.

Advertising