Read plate – Bio-Rad Bio-Plex Pro™ TGF-β Assays User Manual

Page 27

Advertising
background image

8. Read Plate

Bio-Plex Manager

software is recommended for all Bio-Plex Pro

assay

data acquisition and analysis. Instructions for Luminex xPONENT software
are also included. For instructions using other xMAP system software
packages, contact Bio-Rad Technical Support or your regional Bio-Rad
field applications specialist.

Prepare Protocol in Bio-Plex Manager

Software v 6.0 and Higher

The protocol should be prepared in advance so that the plate is read as
soon as the experiment is complete.

A protocol file specifies the analytes used in the reading, the plate wells
to be read, sample information, the values of standards and controls, and
instrument settings.

Bio-Plex Manager software contains protocols for most Bio-Plex

®

assays.

Choose from available protocols or create a new protocol. To create a new
protocol, select File, then New from the main menu. Locate and follow the
steps under Protocol Settings.

1. Click Describe Protocol and enter information about the

assay (optional).

2. Click Select Analytes and create a new panel. Visually confirm the

selected analytes and proceed to step 3.

a.

Click

the

Add Panel button

in the Select Analytes toolbar.

Enter a new panel name. Select Bio-Plex Pro Assay Magnetic
from the assay pull-down menu. If using Bio-Plex Manager version
5.0 or lower, select MagPlex from the assay pull-down menu.

b.

Click

the

Add button. Enter the bead region number and name

for the first analyte. Click Add Continue to repeat for each
analyte in the assay.

25

Advertising