Bio-Rad Bio-Plex Pro™ TGF-β Assays User Manual

Page 34

Advertising
background image

32

Possible Causes

High Intra-Assay CV

Improper pipetting technique

Reagents and assay components
not equilibrated to room
temperature prior to pipetting

Contamination with wash buffer
during wash steps

Slow pipetting of samples and
reagents across the plate

Bio-Plex wash station:
insufficient washing due to
clogged pins

Possible Solutions

Pipet carefully when adding
standards, samples, detection
antibodies, and streptavidin-PE,
especially when using a multichannel
pipet. Use a calibrated pipet. Change
pipet tip after every volume transfer.

All reagents and assay components
should be equilibrated to room
temperature prior to pipetting.


During the wash steps, be careful
not to splash wash buffer from one
well to another. Be sure that the
wells are filtered completely and that
no residual volume remains. Ensure
that the microplate shaker setting is
not too high. Reduce the microplate
shaker speed to minimize splashing.

Sample pipetting across the entire
plate should take less than 4 min.
Reagent pipetting across the entire
plate should take less than 1 min.

Clean dispensing pins with the thicker
of the 2 cleaning needles provided
with washer. Perform regular rinses to
minimize salt buildup.

Advertising