Reacquire data, Data analysis: removing outliers, Previous versions of bio-plex manager software – Bio-Rad Bio-Plex Pro™ TGF-β Assays User Manual

Page 31

Advertising
background image

29

Reacquire Data

It is possible to acquire data from a well or plate a second time using the
Rerun/Recovery mode located below Start in the Run Protocol step.
Any previous data will be overwritten.

1. Check the wells from which data will be reacquired.

2. Remove the buffer with the wash method of choice.

3. Add 100 µl assay buffer to each well. Cover the filter plate with a

new sheet of sealing tape. Shake the plate at 850 ± 50 rpm for
30 sec. Slowly remove the sealing tape before placing the plate on
the plate reader.

4. Repeat the Acquire Data steps to reacquire data. The data acquired
should be similar to those acquired initially; however, the acquisition
time will be extended because the wells have fewer beads.

Data Analysis: Removing Outliers

Outliers are identified as standard data points that do not meet accuracy
or precision requirements and should be considered invalid when
performing curve fitting. As such, they should be removed to generate a
more realistic and accurate standard curve. This may result in an extended
assay working range and allow quantitation of samples that might
otherwise be considered out of range (OOR).

In Bio-Plex Manager software version 6.0 and higher, outliers can be
automatically removed by selecting the Optimize button in the Standard
Curve window. In Bio-Plex Manager software 6.0 and earlier versions,
outliers can be manually selected in the Report Table. Visit online Help to
learn more about the standard curve optimizer feature and how outliers
are determined.

Previous Versions of Bio-Plex Manager Software

For instructions on using previous versions of Bio-Plex Manager software,
please contact Bio-Rad Technical Support.

Advertising