Section 7
Shutdown, Cleaning, Storage Sequence,
and Data Export

After the run is completed, the instrument will instruct the user to begin

the proper cleaning and storage sequence. To maintain the instrument
in proper working order and to prevent mechanical breakdown,
always follow the short- or long-term storage commands
. Short-term
storage is used when the instrument will be run within 2 days, and
long-term storage is appropriate when the instrument will be used in 2 or
more days.

Prepare System to Clean Lines

Remove sample tubes, install tubes filled with water, and install the
cleaning tray. Press Start to begin the cleaning process. Refer to
Figure 16 for the screen associated with this command

Fig. 16. Preparing the system for cleaning.

Export Data

Once the sample and fraction lines are cleaned, the data can be
exported through a USB drive. Install a USB drive into the USB port
located on the front of the instrument, and press Export Data

The exported data can be viewed on any PC using the Profinia
software package

Short-Term/Long-Term Wash

Choose Short-Term Wash to store for <2 days. Choose Long-Term
Wash to store for >2 days