Bio-Rad ProteinChip Qualification and Calibration Kits User Manual

Page 11

Advertising
background image

8

3.

In the Calibration protocol name field, enter

OQ Mass Cal.

4.

Select [Arg8]-Vasopressin (1084.247) from the drop-down list, and click Add
(keep the default values of 3% of expected peaks and minimum signal/noise
of 5).

5.

Select Dynorphin A (2147.5) from the drop-down list, and click Add (keep the
default value of 3% of expected peaks and signal/noise of 5).

6.

Select Beta-endorphin (3465.0) from the drop-down list, and click Add (keep
the default value at 3% of expected peaks and signal/noise of 5).

7.

Select Arg-insulin (5963.8) from the drop-down list, and click Add (keep the
default value at 3% of expected peaks and signal/noise of 5).

8.

Select Fit all 3 coefficients. Click OK.

10010680A.qxp 8/17/2007 1:58 PM Page 8