Bio-Rad ProteinChip Qualification and Calibration Kits User Manual

Page 21

Advertising
background image

9. Plot all 8 spectra. Using the peak selection tool, select the insulin peak (5.963 kD) in all spectra.

10. Export the data following the instructions in Appendix C and selecting the parameters for

export in the following order:

Array bar code

Spot name

Substance mass

Intensity

Resolution

S/N

11. Paste the data into the ProteinChip SELDI OQ form, data sheet “Test 2 Mass Drift and

Resolution”.

12. The spreadsheet indicates whether the test passed or failed.

13. If the test fails, run the HV conditioning procedure (maintenance procedure 1) and repeat the

test. Contact technical support if the test fails a second time.

18

10010680A.qxp 8/17/2007 1:58 PM Page 18