Bio-Rad ProteinChip Qualification and Calibration Kits User Manual

Page 29

Advertising
background image

9. Zoom in and check that all 7 peaks in all 8 spectra are marked correctly. Failure to do so

may result in incorrect results. Close the plot.

Mass Accuracy — External Calibration Test

Perform this test BEFORE the internal calibration test.

1. Click on the spectrum for spot D.

2. Select Spectra > Calibrate > Apply Calibration Protocol.

3. Select OQ Mass Cal.

4. Click OK. The spectrum flag changes from yellow to green.

5. In the main window, make sure that only the spectrum from spot D is selected. Click the

Copy Calibration button. The spectrum flag on spot D reverts to yellow.

6. Select all 8 spectra and click the Apply Calibration button. (Do not select the first

optimization spectrum.)

7. Export the data following the instructions in Appendix C and selecting the parameters for

export in the following order:

Array bar code

Spot name

Substance mass

Intensity

Resolution

S/N

Peak #

8. Paste the data into the ProteinChip SELDI OQ form, data sheet “Test 4 Mass Accuracy”.

Paste the data into cell M41.

26

Spectrum
flag

10010680A.qxp 8/17/2007 1:59 PM Page 26