Bio-Rad ProteinChip Qualification and Calibration Kits User Manual

Page 18

Advertising
background image

2.4.2 Test 2 — Mass Drift and Resolution 5.96 kD

This test uses the ProteinChip peptide standard array to measure the mass drift and resolution
of insulin. These measurements are compared to a specification, and a pass/fail disposition is
obtained.

Array Type

ProteinChip peptide standard array

Usage Information

Each test uses 1 partition on all 8 spots on the array. The array
can be used up to 20 times (5 times on each of 4 partitions).
Store the array in the original packaging and record usage
information in the usage chart provided on the packaging. Laser
optimization steps use 1 of 15 partitions, and these reads do
not need to be recorded or monitored.

Recommended Testing

Weekly

Frequency

Expected Testing Time

20 min to 1 hr

Pass/Fail Specification

Mass drift: pass if less than 7 Da
Resolution: pass if average is greater than 750

1. Insert the ProteinChip peptide standard array into the instrument. If you are using an

Enterprise Edition instrument, place the array into a cassette, fill the remaining slots in the
cassette with blank or used arrays, and place the cassette into the instrument.

2. In ProteinChip data manager software, click Start and open the Protocol Mode tab.

3. Select Instrument > Preferences. In the Preferences dialog, click “…”. The Select Folder

dialog opens.

4. Select Projects > OQ_####/

Date/Test 2 Mass Drift and Resolution 5.96 kD, where ####

is the serial number of the instrument.

5. Optimize the laser intensity. Run protocol “Test 2 Mass Drift and Resolution 5.96 kD” on any

partition (“n”) of 15 on spot D (for example, 1 of 15, 2 of 15, 3 of 15, etc). Adjust the laser
energy until the height of the insulin peak at 5,963 Da is 100–200 µA. This optimization step
does not need to be repeated on subsequent runs of test 2 unless the peak intensity falls
too low.

15

10010680A.qxp 8/17/2007 1:58 PM Page 15

Advertising