Bio-Rad ProteinChip Qualification and Calibration Kits User Manual

Page 15

Advertising
background image

2.4 OQ Tests

2.4.1 Test 1 –– Detector Sensitivity

This test uses the ProteinChip detector qualification array to measure the signal-to-noise ratio
(S/N) of immunoglobubulin (IgG) at two different concentrations (10 fmol and 140 fmol).
Measurements are compared to a specification, and a pass/fail disposition is obtained.

Array Type

ProteinChip detector qualification array

Usage Information

Use once and discard; do not open until ready to use.

Recommended Testing

Every other week

Frequency

Expected Testing Time

30 min to 1 hr

Pass/Fail Specification

High concentration: pass if S/N is greater than 1,100
Low concentration: pass if S/N is greater than 5

1. Insert the ProteinChip detector qualification array into the instrument. If you are using an

Enterprise Edition instrument, place the array into any location within the cassette, fill the
remaining slots in the cassette with blank or used arrays, and place the cassette into the
instrument.

2. In ProteinChip data manager software, click Start and open the Protocol Mode tab.

3. Select Instrument > Preferences. In the Preferences dialog, click “…”. The Select Folder

dialog opens.

4. Select Projects > OQ_####/

Date/Test 1 Detector Sensitivity, where #### is the serial

number of the instrument.

5. Run protocol “Test 1 Detector Sensitivity ” on partition 1 of 1 for all 8 spots on the array. Do

not change any values.

12

10010680A.qxp 8/17/2007 1:58 PM Page 12