Filter size, Log scale output – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 45

FILTER
SIZE
Calls up the numeric entry screen to select the number of samples to use for
each analog output update. The saved sample represents a sliding window and
serves as a smoothing function for the analog output signal. The analog output
level is updated every sample according to the samples per second setting in the
Main Menu.
•
If the filter size is greater than one, the sample window contains the
current sample plus the most recent samples.
•
If the filter size is set to one, only the current sample is used by the
analog output system.
The minimum/maximum/average setting affects how the data is processed. See
Min/Max/Avg on page 44.
LOG
SCALE
OUTPUT
Selects either linear or log scale for the analog output voltage.
•
When selected, the output will track a logarithmic curve with a zero
field level being zero volts out and a full scale field level being five volts
out.
•
When deselected, the endpoints will be the same but the analog
voltage out to field level ratio will be linear.
Menu System
| 45