LED
S
ETUP
Description:
Configures the FM5004 LED array. This command is
used to select probes for display or analog output
functions and to enable and disable probe axes.
Separate commands must be sent to update LEDs for
each channel.
Syntax:
CLED,m,xxxxx
Parameters:
Channel number (m):
• 0 = all channels set the same
• 1 = channel 1
• 2 = channel 2
• 3 = channel 3
• 4 = channel 4
LED configuration (xxxxx):
• This is a five-bit binary number.
A one corresponds to an LED illuminating, and a
zero corresponds to an LED going dark.
The order of the bits is: display, out, x, y, z.
Example:
To enable channel 1 for display output with all three
axes enabled:
CLED,1,10111<LF>
The LEDs will change accordingly and the FM5004 will
return:
CLED,1,10111<CR><LF>