Alarm upper value – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 69

A
LARM
U
PPER
V
ALUE
Description:
Sets the upper levels for the line 1, line 2,
4-probe/3-axis, and high temperature alarms. Each of
the alarm levels must be set using individual
commands.
For an alarm to operate, that alarm must be enabled.
The data entered can contain up to five numbers and a
decimal point.
See the description for each alarm type for the
permissible range. See Alarm Upper Enable on
page 68 to enable an alarm.
Syntax:
AUPV,m,xxxxxx
Parameters:
Alarm Type (m):
• 0 = high temperature alarm,
range: 32
°
F–211
°
F
• 1 = line 1 alarm,
range: 0—max of probe
• 2 = line 2 alarm,
range: 0—max of probe
• 4 = 4-probe/3-axis alarm,
range: 0—max of lowest range probe
Alarm level (xxxxxx)
• xxxxxx = alarm value: 0–1999.9
Example:
• To set the line 1 upper alarm value to 150.5:
AUPV,1,150.5<LF >
• To verify the line 1 upper alarm value:
AUPV,1?<LF >
The FM5004 will return:
AUPV,1,150.5<CR><LF >
Remote Operation
| 69
Archived 10/19/10