Alarm lower value – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 67

A
LARM
L
OWER
V
ALUE
Description:
Sets the lower levels for the line 1, line 2,
4-probe/3-axis, and high temperature alarms. Each of
the alarm levels must be set using individual
commands.
For an alarm to operate, that alarm must be enabled.
The data entered can contain up to five numbers and a
decimal point.
See the description for each alarm type for the
permissible range. See Alarm Lower Enable on
page 66 to enable an alarm.
Syntax:
ALOV,m,xxxxxx
Parameters:
Alarm Type (m):
• 0 = low temperature alarm,
range: 32
°
F–211
°
F
• 1 = line 1 alarm,
range: 0—max of probe
• 2 = line 2 alarm,
range: 0—max of probe
• 4 = 4-probe/3-axis alarm,
range: 0—max of lowest range probe
Alarm level (xxxxxx)
• xxxxxx = alarm value: 0–1999.9
Example:
• To set the line 1 lower alarm value to 150.5:
ALOV,1,150.5<LF>
• To verify the line 1 lower alarm value:
ALOV,1?<LF>
The FM5004 will return:
ALOV,1,150.5<CR><LF>
Remote Operation
| 67