Line 1 alarm / line 2 alarm – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 49

• Lower Value:
Calls the numeric entry screen to enter the lower
measured field value below which the alarm will
activate.
• Latched:
The alarm will sound continuously when an enabled
upper or lower value is crossed, even if the measured
field returns to within the normal field level.
To temporarily disable the alarm, select Alarm Silence
from the Main Menu or power the FM5004 on and off.
LINE 1
ALARM
/
LINE 2
ALARM
Contains the items associated with setting up the alarm for use while in 2-line
display mode. There are separate alarm setup menus for each line in 2-line
display mode.
Using the alarms while in 2-line display mode allows the use of two
different upper and lower alarm values for different probes.
• Upper Enable:
The alarm will activate when the measured field is
greater than the upper alarm value.
• Upper Value:
Calls the numeric entry screen to enter the upper
measured field value above which the alarm will
activate.
Menu System
| 49
Archived 10/19/10