Analog output range – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 71

A
NALOG
O
UTPUT
R
ANGE
Description:
Sets the analog output full scale range to correspond to
the maximum of range 1 through range 4 of the analog
output selected probes.
All analog output selected probes must use the same
units of measurement and have the same ranges.
Syntax:
ANAR,x
Parameters:
Range value (x):
• 1 = range 1 upper limit corresponds to full scale
analog output voltage
• 2 = range 2 upper limit corresponds to full scale
analog output voltage
• 3 = range 3 upper limit corresponds to full scale
analog output voltage
• 4 = range 4 upper limit corresponds to full scale
analog output voltage
Example:
• To set the analog output full scale voltage to
correspond to range 2 of the analog output selected
probes:
ANAR,2<LF>
• To verify the analog output full scale voltage
corresponds to range 2 of the analog output
selected probes:
ANAR?<LF>
The FM5004 will return:
ANAR,2<CR><LF>
Remote Operation
| 71