• Upper Enable:
The alarm will activate when temperature of any active
probe rises above the upper alarm value.
• Upper Value:
Calls the numeric entry screen to enter the upper
temperature level above which the alarm will activate.
• Lower Enable:
The alarm will activate when a probe temperature falls
below the lower alarm value.
• Lower Value:
Calls the numeric entry screen to enter the lower
temperature level below which the alarm will activate.
• Latched:
The alarm will sound continuously when an enabled
upper or lower temperature level is crossed, even if the
measured temperature returns to within the normal
field level.
To temporarily disable the alarm, select Alarm Silence
from the Main Menu or power the FM5004 on and off.