Alarm lower enable – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 66

A
LARM
L
OWER
E
NABLE
Description:
Enables the line 1, line 2, 4-probe/3-axis, battery, or
temperature lower limit alarms. Each of these alarms
must be enabled and disabled separately.
In the case of the low battery alarm, the level is not
adjustable. See the Battery on page 30 for a
description of the low battery alarm operation. See
Alarm Lower Value on page 67 to set the alarm lower
values.
Syntax:
ALOE,m,x
Parameters:
Alarm Type (m):
• 0 = low temperature alarm enable
• 1 = line 1 lower alarm enable
• 2 = line 2 lower alarm enable
• 4 = 4-probe/3-axis lower alarm enable
• 5 = low battery alarm enable
Numeric value (x):
• 0 = lower alarm disabled
• 1 = lower alarm enabled
Example:
• To enable the line 1 lower alarm:
ALOE,1,1<LF>
• To verify the line 1 lower alarm status:
ALOE,1?<LF>
The FM5004 will return:
ALOE,1,1<CR><LF>
66 |
Remote Operation
Archived 10/19/10