Temp alarm setup – ETS-Lindgren FM5004 Field Monitor (Archived) User Manual
Page 50

• Lower Enable:
The alarm will activate when the measured field falls
below the lower alarm value.
• Lower Value:
Calls the numeric entry screen to enter the lower
measured field value below which the alarm will
activate.
• Latched:
The alarm will sound continuously when an enabled
upper or lower value is crossed, even if the measured
field returns to within the normal field level.
To temporarily disable the alarm, select Alarm Silence
from the Main Menu or power the FM5004 on and off.
TEMP
ALARM
SETUP
Contains the items
associated with
setting up the alarm
for an out of range
temperature
condition.
The temperature alarm will become inactive when the sample rate is set above
10 samples per second. However, taking a temperature measurement from
either the front panel or by a remote temperature measurement command will
cause an alarm if the alarm is enabled and the measurement causes an alarm
condition.