L
OCAL
M
ODE
Description:
Returns the FM5004 to local mode when it is in
remote mode. Local mode allows the operator to use
the front panel buttons.
Syntax:
LOCL,x
Parameters:
Local mode enable (x):
• 0 = disabled; stays in remote mode
• 1 = local mode enabled
Example:
To return the FM5004 to local mode:
LOCL,1<LF>
The front panel buttons will become operational and
the FM5004 will return:
LOCL,1<CR><LF>
L
OG
S
CALE
Description:
Enables either a logarithmic or linear scale for the
analog output voltage.
Syntax:
LOGS,x
Parameters:
Log scale enable (x):
• 0 = log scale disabled (linear scale enabled)
• 1 = log scale enabled
Example:
To set the analog output voltage to log scale:
LOGS,1<LF>
The scale will change to logarithmic scale and the
FM5004 will return:
LOGS,1<CR><LF>