Page 13

Advertising
background image

Manual No. 085-3435-07

12/45

IGB/IGX Series (Overseas Specifications) Service Manual

2.2 C1 Mode (Country No., Scale No., Zero point, Span adjustment)

2.2.1 Country No. Table

Country Name

USA 1 EU AUS ASIA USA 2

JPN

IDV

SET

Country Code

0

1

2

3

4

5

6

7

8

9

10

99

ADRS POS WGT

Function

Default Data (Change not allowed)

112 A H

Start

width

2 2

2 2 2 2

2

2

2

2 4 -

B H

Stable/Re-stable

count

3 3

3 3 3

3

3

3

3

3

3 -

113 A H

Stable/Re-stable

width

4 4

4 4 4 4

4

4

4

4 4 -

B H

Re-stable operation start width

4 4

4 4 4 4

4

4

4

4

4 -

114

A

1

Zero point mark

2

Over-scale

display

4

Display less than true zero

8

Decimal point display

2 8

0 8 2 8

0

8

4

0

5 -

B 1

Over-scale

range

2

Tare

subtraction

4

Tare clear with ZERO key

8

Zero

suppress

0 6

6 0 0 6

6

0

0

6

0 -

115

A

1

Key-in tare subtraction

2

Zero

tracking

4

Micro weight follow-up

8

Unstable

width

0 0

0 0 0 0

0

0

0

0

4 -

B

1

PLU display (IGB only)

2

Stable display (IWB only)

4

Reserved

8

Reserved

0 6

6 6 4 6

6

6

5

5

5 -

Note

1. By setting the country code, weighing conditions will function based on the default data from

address 112 to 115.

2. The data is set to meet certified conditions for each country, so there is no need to change the

data.

3. Position “A” indicates the upper position of one-byte data, and “B” indicates the lower position.
4. As for Weight, a function is selected with “1”, “2”, “4”, and “8” in bit unit, and “H” indicates

Hexadecimal data.

5. Individual setting value “99” cannot be entered. (Displayed only when settings are changed in C7

mode)

6. “USA 2” code is settings to allow changing between “lb” and “kg” at anytime even in cases other

than a stable condition at zero point. (“USA 1” can change between “lb” and “kg” only when the
condition is stabilized at zero point)

When using this function, a maximum of 0.2e error may occur in weighing immediately after the

change. This error is calibrated by zero point adjustment (or zero tracking), and an accurate
weighing is guaranteed.
When selecting the weighing capacity exclusively for either “lb” or “kg”, use “USA 1”.

Reference

Reference

Refer to “Weighing Condition Data Table” for detailed function selection.