Manual No. 085-3435-07

15/45

IGB/IGX Series (Overseas Specifications) Service Manual

Operation Display

6. Span Adjustment
Place the selected weight on the weighing platter, then press

TARE(-).

[Example] 60kg Scale: Weight selection: “kg” unit and “1/1”

Place a 60kg weight, 30000 + Zero point 20000 =
50000 counts

In case of “kg” unit “1/2” for a weight selection, place
a 30kg weight, then 15000 + Zero point 20000 =
35000 counts

One graduation equals 5 counts.

Processing is automatically executed in the following order:
[Calib] (Calibration) Computing process executing
[C- OK] (Calibration OK) Computing result succeeded
[No. of adjusted counts] Computing result

[C-Err] will be displayed, indicating a computing result error
when span adjustment is performed without placing a weight.

Place a weight and press TARE(-).

¾

¾

(Calibration error)

7. Saving Data
To save Country No., Scale No., Zero point, and Span data in
E2ROM.

Remove the weight from the weighing platter.
Remove the seal covering the opening on the rear case of

display unit.

Insert a thin rod such as the inner shaft of a ball-point pen

into the hole, and push the Memory Switch located on the
main board.

When writing has finished normally, [S-OK] will be

displayed.

To release the above status, press COLUMN.

Then, the display will return to A/D data.



8. Finishing Procedure

Press ON/OFF to finish this procedure.