IGB/IGX Series (Overseas Specifications) Service Manual

2.8 C7 Mode: Weighing Condition Setting

Note

Setting Country No. and Scale No. determine the weighing conditions for securing certified
specifications of the country.
Individual contents can be changed in this mode.
Specifications must be selected when used overseas.

Operation Display

1. Starting Test Mode

Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.

2. Entering C7 Mode

Press ZERO(+) six times.

Or, press TARE(-) three times.

3. Press SET.
To change data:
Select the desired data by pressing ZERO(+) to increase,

or TARE(-) to decrease the address.


To fix the data or advance the address:
Press

SET.

3-digit address + A (Upper data) - Data


3-digit address + B (Lower data) - Data

4. After changing data
Push the Memory Switch (SW1) located on the main

board.

(This does not work when there was no data change)

5. Press

COLUMN twice to return to Test Mode.

Reference

Reference

Addresses for weighing condition settings are from 110 to 115.
Refer to the next page.