IGB/IGX Series (Overseas Specifications) Service Manual

Weighing Condition Setting Table

Address Weight

Item

Data

110A

1

Decimal point position

0: 0

1: 0.0

2

2: 0.00

3: 0.000

4

Minimum graduation

0: 1 (1-2 change in Dual range/ Multi Interval)
4: 2 (2-5-10 change in Dual range/ Multi Interval)

8

8: 5 (5-10-20 change in Dual range/ Multi Interval)
C: Invalid

110B -

Accuracy

(Resolution)

0:

1/500

1: 1/600

2: 1/750

3: 1/1000

4: 1/1200

5: 1/1500

6: 1/2000

7: 1/2500

8: 1/3000

9: 1/4000

A: 1/5000

B: 1/6000

C: 1/7500

D: 1/10000 E: 1/12000 F: 1/15000

1

2

4

Changing method

0: Single range LED stable display only (IGX)
1: 1/3000 Multi interval Stabe/Upper/Lower display (IGB)
2: Single range Upper/Lower display (IGX)
3: 1/3000 Multi interval Stabe/Upper/Lower display (IGB)
4: Single range (Body weight specification)
5: 1/3000 Dual range (A/B range change)
6: Single range (lb/kg change)
7: 1/3000 Multi interval (lb/kg change)

111A

8

1

Weight unit

(Valid only for
communications)

<111A> <111B>
0 + 0: kg (kg - lb)
8 + 0: g
0 + 1: lb
8 + 1: oz

2

4

111B

8

Filter setting



Do not change the data.

(Cut off)

(Notch) (Output rate)

0: 0.66Hz

2.50Hz

400ms

2: 0.84Hz

3.20Hz

312ms

4: 1.05Hz

4.00Hz

250ms

6: 1.31Hz

5.00Hz

200ms

8: 1.68Hz

6.40Hz

156ms

A: 2.10Hz

8.00Hz

125ms

C: 2.62Hz

10.00Hz

100ms

E: 3.35Hz

12.80Hz

78ms