3 operation procedure – Rice Lake Ishida IGX & IGB Series - Service Manual User Manual
Page 15

Manual No. 085-3435-07
14/45
IGB/IGX Series (Overseas Specifications) Service Manual
2.2.3 Operation Procedure
Operation Display
1. Starting Test Mode (C1)
Press ON/OFF.
Press and hold ZERO(+) while all segments are lit.
Press SET.
2. Setting Country Number
To increase the number, press ZERO(+).
To decrease the number, press TARE(-).
Press SET.
[Example] Asia = 3
3. Setting Scale Number
When setting “60kg”, select “08”.
Press SET.
4. Selecting Weight at Span Adjustment
The following weight can be selected:
kg: 1/1.0→1/2.0→1/2.5→1/5.0 of weighing capacity
lb: 1/1.0→1/2.0→1/2.5→1/5.0 of weighing capacity
Pressing ZERO(+) selects the next item in the above
sequence, and pressing TARE(-) selects the previous item.
kg unit: Same weight is used as weighing
capacity.
[Example] Selecting “kg” unit and “1/2.0” for weighing
capacity “150kg” will require the total 75kg
weight.
kg unit: Half weight is used of weighing
capacity.
To change between “lb” and “kg”, press ZERO(+) and
TARE(-) at the same time.
lb unit: Same weight is used as weighing
capacity.
5. Zero Point Adjustment
After selecting the weight amount, ensure that nothing is
placed on the weighing platter, then press SET.
The previously set zero point A/D data will be displayed.
Zero point indicator W will light up.
Press ZERO(+) to set “20000” counts forcibly.
Zero point indicator W will go off.