Specifications, 10 - specifications – Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual

Page 103

Advertising
background image

10

99

Specifications

ECG

The ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and
Alarms specified ANSI/AAMI EC13-1992, except for Standardizing Voltage (section
3.2.9.9). The channel also meets the American National Standard, Safe Current Limits for
Electromedical Apparatus (ANSI/AAMI ES1-1993).

Characteristic

Specification

Connector

AAMI 6 pin or Hewlett-Packard compatible 12-pin style connector (optional). (See the
illustration at the bottom of this table.)

Selectable Leads

I, II, III, aVR, aVL, aVF, V

Lead Fault Indicator

LA, LL, RA, RL, C, multiple

ECG Size (sensitivity) in mV/cm

4, 2, 1, 0.5, 0.2

Display Sweep Speeds

12.5, 25, and 50 mm/sec

QRS Tone Volume

High, Low, Medium, Off

QRS Tone Frequency

900 Hz. for Propaq Encore without Expansion Module, 665 Hertz when equipped with
SpO2 but SpO2 not being monitored; variable pitch with SpO

2

option and SpO

2

being

monitored

Freeze Buffer

3.9 seconds at 25 mm/sec

Bandwidth

0.5 to 40 Hz in Monitor Mode; 0.05 to 40 Hz in Extended Mode (see Real-Time ECG
Analog/Defib Sync specification).

Sample Rate

364 Hz

Input Protection

Electrosurgery and defibrillator protected when used with specified ECG cables. All
models also include electrosurgery interference suppression.

Lead Fail Sense Current

50 nA dc for active leads
100-200 nA dc for driven lead, depending on number of electrodes attached

Tall T-wave Rejection

Meets and exceeds AAMI (USA) EC13-1992, section 3.1.2.1.c, for 1.2 mV T-wave and
1 mV QRS using AAMI test waveform.

Common Mode Rejection

<1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source impedance, input
unbalanced, FILTER function OFF
<0.1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source impedance, input
unbalanced, FILTER function ON

Input Impedance

>2.5 M

Ω

differential @ 60 Hz

Input Range (ac)

10 mV peak to peak

Input Range (dc)

Up to ± 300 mV

System Noise

≤30 µV peak-to-peak, R.T.I., with all inputs = 47K in parallel with .047 µF.