Care and maintenance

Welch Allyn Propaq Encore Vital Signs Monitor

Maintenance

Service interval recommendations

At the intervals recommended below, verify the Propaq Encore for proper operation of all
channels and internal circuitry. Such checks and verifications should only be carried out by
a qualified biomedical service person.

Other Propaq Encore service information, including calibration procedures, is described in
the Propaq Encore Service Manual (P/N 810-0696-XX). Refer to it for more information.

Use the following intervals for a guideline. Service may be needed more often in extreme
environments (heat, cold, dust, etc.).

d.

The mainstream CO

2

sensor may also be disinfected with Wex-cide. Follow the disinfectant manufacturer’s instructions. Do not leave

Wex-cide on sensor longer than 30 minutes. Thoroughly clean off residue with water-dampened cloth. Prolonged exposure of the sensor to
Wex-cide will damage the sensor.

Caution Do not autoclave this product or its accessories. Do not immerse the
monitor in liquid when cleaning. Do not immerse accessories in liquid when
cleaning unless the accessory manufacturer’s cleaning instructions instruct you
to do so.

Recommended interval

Service action

Six months to two years

Complete functional verification; see Propaq Encore Service Manual
Inspect the Propaq Encore for mechanical and functional damage
Inspect safety labels for legibility
Inspect the side panel fuse for compliance to specified rating
Verify that visual and acoustic alarms are functioning properly
Test patient leakage current according to IEC 601-1/1988
Test patient leakage current with mains voltage on patient-applied parts according to IEC
601-1/1988: limit 50

μA

a

a.

NOTE: The leakage current should never exceed the 50µA limit. The data should be recorded in an equipment log. If the device is not
functioning properly or fails any of the above tests, do not attempt to repair the device. Please return the device to the manufacturer
or to your distributor for any required repairs.

Minimum every three years

Check battery capacity