20

Getting started

Welch Allyn Propaq Encore Vital Signs Monitor

Monitor setup

Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu.

STATSCALE

Automatically readjusts all waveform scales.

ALARMS

Allows access to the Alarms menu.

WAVE SEL

Allows you to turn on and off desired waveforms or NIBP numerics
for display.

TRENDS

Allows access to the Trend settings and display.

MORE

Displays the next setup menu and the following status window:

NEXT

Selects the next setting in the status window.

CHANGE

Changes the currently selected display setting. (Pressing CHANGE
at PATIENT MODE allows you to choose between Adult, Pediatric,
and Neonatal in a Patient Mode window.)

PRINTER

Allows access to the Printer Menu.

MORE

Allows access to the Time/Day window.

CURRENT
SOURCE

When the selected HR/PR source is no longer available, the current
source is the active source with highest priority. The RR/BR source
cannot be manually selected. It will always be CO

2

if CO

2

is active.

Otherwise, it will be ECG/RESP.

SELECTED
SOURCE

The user-selected HR/PR source is displayed along with the HR/PR
source currently being used by the monitor.

SWEEP (mm/s) The selectable sweep speeds for HR/PR are 12.5, 25, and 50 mm/

sec. The sweep speeds for RR/BR are 3.13, 6.25, and 12.5 mm/sec.

ALARM TONE

Sets the Alarm Tone volume to HIGH, MEDIUM, or LOW.

STATSCALE

MORE

WAVE SEL

TRENDS

ALARMS

mmHg

NEXT

MORE

WAVE SEL PRINTER

CHANGE

MCO

2

BR

SpO

2

PA

SETUP

HR/PR

HR/PR RR/BR

CURRENT SOURCE : ECG MCO2
SELECTED SOURCE : ART ---
SWEEP (mm/s) : 12.5 3.13
ALARM TONE : HIGH
HR/PR TONE : LOW
PATIENT MODE :

ADULT

35/ 18

( 85)

35 12

92