46

Patient monitoring

Welch Allyn Propaq Encore Vital Signs Monitor

START/STOP Starts and stops NIBP measurements. Any time the Propaq Encore is

taking a noninvasive pressure measurement, the

START

button

changes to

STOP

so you can stop the measurement in progress. This

button initiates the same action as the

START/STOP

button at the left

side of the screen. Pressing

STOP

will automatically vent the cuff.

AUTO/MAN

This button switches the mode between Automatic or Manual Mode.
The Manual Mode is the default unless you change it by
reprogramming your Propaq. Measurements can be taken at intervals
of 1, 2, 3, 5, 10, 15, 30, and 60 minutes. Press

START

to initiate a

measurement.

INTERVAL

Selects the interval at which NIBP measurements are automatically
taken. The interval you select, ranging from one minute to 60 minutes,
is shown on the display next to the word TIME.

TURBOCUF

Automatically starts NIBP measurements and continues to take as
many measurements as possible within five minutes.

SMARTCUF

Enables or disables the Smartcuf motion artifact filter. NIBP
measurements can still be taken when Smartcuf is off. Artifact may
interfere with the accuracy of NIBP measurements with Smartcuf off.

STOP

SMARTCUF

INTERVAL TURBOCUF

AUTO/MAN

ART

HR

T1

NIBP

18:46:00

18:45:28

III

1

mV

cm

MON

NIBP ADULT

MODE : TURBOCUF
TIME : 4:21
SMARTCUF : ON

102.3

F

S 122

( 87)

85

109

72

122

58

80

85

50 100 150 200

85

122

58

NIBP

99

SpO2

mmHg

mmHg

122

58

85

This symbol indicates the NIBP reading was taken in
the presence of high motion artifact while
monitoring ECG. Artifact can affect accuracy. To help
reduce artifact, see page 45.

NIBP

122/

58

( 85)

18:45

Manometer Bar.

Systolic, diastolic, and

mean values are indicated

above the manometer as

small triangles.

Time of last measurement

Last measurement

Measurement mode

Alarm Limits Bell

NIBP values and units of measure

rotate in and out of this window

for up to one hour (International

English only)

NIBP values are
displayed in one of
these windows if
vacant. After 16
minutes, the display
changes to that shown
below for 44 minutes.

Time remaining in Turbocuf

Mode or Auto Mode interval