Sidestream co, Menu and status window – Welch Allyn 206EL Propaq Encore Vital Signs Monitor - User Manual User Manual

Page 64

Advertising
background image

60

Patient monitoring

Welch Allyn Propaq Encore Vital Signs Monitor

Sidestream CO

2

menu and status window

When sidestream CO

2

is active, press MORE from the first CO

2

menu to access the

SCO

2

menu and the SCO

2

status window:

SOURCE

Changes between mainstream CO

2

and sidestream CO

2

monitoring

(if both options are installed), or disables CO

2

monitoring

Setting CO2 SOURCE to OFF allows you to disable CO

2

monitoring

without removing the sensor. When the CO2 SOURCE is set to OFF,
the Propaq displays OFF for CO

2

numerics.

SWEEP SPEED The CO

2

and RESP display sweep speed (set by the mm/s button)

BAROMETER

The ambient barometric pressure

Note

To change CO

2

units, use the Settings Menu under the Service Menus. Changing

these units will clear CO

2

trends.

GAS COMP

Selects the measurement compensation for CO

2

measurements

RESPONSE

Sets the response time for CO

2

measurement (NORMAL, FAST, OR

SLOW)

SOURCE

Changes between mainstream CO

2

and sidestream CO

2

monitoring

(if both options are installed), or disables CO

2

monitoring

Setting CO2 SOURCE to OFF allows you to disable CO

2

monitoring

without removing the sensor. When the CO2 SOURCE is set to OFF,
the Propaq displays OFF for CO

2

numerics.

SWEEP SPEED The CO

2

and RESP display sweep speed (set by the mm/s button)

BAROMETER

The ambient barometric pressure

FLOWRATE

Sets the sampling flow rate to either 90 or 175 ml/min. You can
change the flow rate while sidestream CO

2

is active

Note

To change CO

2

units, use the Settings Menu under the Service Menus. Changing

these units will clear CO

2

trends.

mmHg

GAS COMP

PREV MENU

RESPONSE

SCO

2

BR

SpO

2

PA

CO

2

GAS COMPENSATION: OFF
RESPONSE : NORMAL
CO2 SOURCE : SIDESTREAM
SWEEP SPEED : 6.25 mm/s
BAROMETER : 762.0 mmHg

35/ 18

( 85)

35 12

92

SOURCE

FLOWRATE : 175

ml/min

FLOWRATE

Advertising