Specifications

105

Temperature

Characteristic

Specification

Range

0° to +50°C; 32° to +122°F

Displays

T1, T2, and

Δ

T

Probes

Compatible with YSI Series 400 and 700 probes. HP side panel
only compatible with YSI 400 and has HP connector.

Units

°C and °F selectable

Channel Accuracy

Temperature Range

Tolerance

0° to +10°C
>10° to +50°C

+32° to +50°F
>50° to +122°F

± 0.2°C
± 0.1°C

± 0.4°F
± 0.2°F

Resolution

0.1°C or °F

Temperature Alarm
Limit Range (T1, T2)

32.0° to 122.0°F
0° to +50.0°C

Electrosurgery
interference
suppression

Included in all models: 202EL, 204EL, 206EL