Alarms and limits

Welch Allyn Propaq Encore Vital Signs Monitor

Setting alarm limits

Adjust limits with STAT SET

When you want to quickly set all alarm limits, the Propaq Encore can calculate new alarm
limits using the patient’s current values. Press the STAT SET button in the Alarms Menu
(SETUP, ALARMS, STAT SET). The Propaq Encore activates all alarms and calculates the
limits for all monitored vital signs, except apnea delay. The monitor performs a
mathematical function (addition, subtraction, or multiplication) on the current value of the
vital signs to arrive at the new limits. The formulas for these calculations are shown in the
table.

Vital sign

If the patient’s vital sign
value is

Then calculated new lower
limit is

Then calculated

a

new upper

limit is

a.

New alarm limits calculated by STAT SET cannot be outside the allowable alarm limit range. If a new limit is calculated to be above
or below the allowable alarm limit range, it defaults to the maximum or minimum alarm limit allowed for that vital sign.

Heart Rate

HR

≤ 99

100-250
HR

≥ 251

HR x 0.8
HR - 20
Unchanged

HR x 1.2
HR + 20
250

Pulse Rate

PR

≤ 99

PR

≥ 100

PR x 0.8
PR - 20

PR x 1.2
PR + 20

Invasive
Pressure

Inv Prs

≤ 25

26 - 99
Inv Prs

≥ 100

Inv. Pressure - 5
Inv. Pressure x 0.8
Inv. Pressure - 20

Inv. Pressure + 5
Inv. Pressure x 1.2
Inv. Pressure + 20

NIBP

NIBP

≤ 25

26 - 99
NIBP

≥ 100

NIBP - 5
NIBP x 0.8
NIBP - 20

NIBP + 5
NIBP x 1.2
NIBP + 20

Respiration Rate/
Breath Rate

RR/BR

≤ 25

26 - 99
RR/BR

≥ 100

RR/BR - 5
RR/BR x 0.8
RR/BR - 20

RR/BR + 5
RR/BR x 1.2
RR/BR + 20

Temperature

Temp

≥ 0°C

Temp - 0.5

Temp + 0.5

SpO

2

SpO

2

≥ 0%

SpO

2

- 5

(min. limit 50%)

100% (adult and pediatric
mode)
SpO

2

+ 5 (neonate mode)

ETCO

2

ETCO

2

≥ 0 mmHg

ETCO

2

≥ 2.0 (% or kPa)

ETCO

2

- 5 mmHg

(min. 15 mmHg)
ETCO

2

- 0.7 (% or kPa) (min. 2.0%

or 2.0 kPa)

ETCO

2

+ 10 mmHg

ETCO

2

+ 1.4 (% or kPa)

INCO

2

INCO

2

≥ 0 mmHg

INCO

2

≥ 0 (% or kPa)

Not affected by STAT SET

INCO

2

+ 5 mmHg

INCO

2

+ 0.7 (% or kPa)

Apnea Delay

Not affected by STAT SET