Real-time ECG analog/defib sync

Special cables are required to interface the defib sync connector to the Physio-Control
LIFEPAK 5 or LIFEPAK 6s defibrillator.

The sync and real-time ECG outputs do not operate during In-service mode.

Signal

Specification

Sync Output

0 to 5V pulse, 100 ± 5ms wide, starts within 35 ms after peak of R-wave. 15 mA short
circuit current.

Real-time ECG Output

Range = ± 6V minimum, centered about 0V, Gain = 1000X, noninverting for lead II,
inverting for all other leads, delay <3 msec, 0.05-100 Hz, going to -5.9V ± 5% during
ECG lead fail. V lead has no Real-Time analog output.

Marker Input (Defib Sync
only)

Normally 0V in, a pulse either + or -3 to ± 15V for 10-70 ms puts a marker in ECG
trace. ~ 5 k

Ω input res.

Shield

Common terminal for other signals

N/C

N/C

GROUND

MARKER IN

SYNC
OUT

N/C

Real-time ECG Output connector

(side-panel view)

Defib Sync connector

(side-panel view)

ECG

N/C

GROUND

SHIELD

N/C

SHIELD