Patient monitoring

Welch Allyn Propaq Encore Vital Signs Monitor

Connect to an intubated patient

1.

Connect the gas sampling elbow and elbow connector into the patient’s breathing
circuit according to the manufacturer’s instructions.

CO

2

messages

Mainstream CO

2

messages

Equipment messages for the mainstream CO

2

option can appear on the display in an

equipment alert window and in numeric zones.

If a sensor is damaged, contact Welch Allyn’ Technical Services Department for
information on sensor service options.

WARNING The exhaust port for sidestream CO

2

is an output for the expired

gases from the patient and any connected breathing apparatus. The exhaust port
is intended only for connection to gas collection equipment such as gas
scavenger devices. Do not allow any other connection to the exhaust port.

Note

If you use a gas scavenging system with sidestream CO

2

, be sure to install it

according to the manufacturer’s instructions. The scavenging system should
comply with ISO 8835-3:1997 (E).

Note

If you are using a watertrap and sample line instead of the one-piece CO2 gas
sampling system, perform steps above with the following changes: attach the
watertrap to the monitor, then attach the sample line to the watertrap.

Exhaust Port

CO2 Watertrap Adapter

Sample Line

Elbow

Connector

WARNING - Do not connect sample line
or patient input to exhaust port.