Patient monitoring

Welch Allyn Propaq Encore Vital Signs Monitor

How to set up the CO

2

channel and alarm limits

After you connect the mainstream CO

2

sensor or sidestream CO

2

watertrap adapter,

follow these steps to set up the CO

2

channel and alarm limits.

1.

Press SpO2/CO2, CO2 to display the first CO

2

Menu.

2.

Press RANGE until you see the desired waveform scale range on the Propaq screen.
The range choices are shown below.

3.

Press mm/s to select either 3.13, 6.25 or 12.5 mm/sec. The default setting is 6.25
mm/sec. This sweep speed also applies to RESP.

4.

Press MORE to view the CO

2

status window.

5.

If either O

2

or N

2

O is being administered to a patient, press GAS COMP to set the

proper gas compensation as listed below. If no other gas is being administered, set
GAS COMPENSATION to OFF. (OFF is the default setting.)

Note

When disconnecting the airway adapter from the ventilator circuit, always detach
the CO

2

sensor from the airway adapter before removing the airway adapter from

the ventilator circuit.

Note

After you connect a mainstream CO

2

sensor or sidestream CO

2

watertrap

adapter, the Propaq displays the waveform briefly without a value range. It
displays WARM UP (for mainstream) or START UP (for sidestream) in the CO

2

numerics window. After about 20 seconds, the Propaq displays the CO

2

measurement and waveform range.

Units

Waveform Scale Range

mmHg

0-100

0-60 (default
setting)

0-30

kPa

0-14

0-8

0-4

%

0-14

0-8

0-4

Note

If an inspired value is displayed indicating patient rebreathing (non-zero IN

CO

2

),

check the patient breathing circuit for proper function. Verify the sensor calibration
against room air. If the Propaq continues to display inspired values, return the
sensor to Welch Allyn for service.

Gas Administration Level/
GAS COMP Setting

ETCO

2

or INCO

2

Value

OFF

CO

2

value = actual

CO

2

value

O

2

> 50%, No N

2

O

CO

2

value = actual CO

2

value x 1.03

N

2

O > 50%

CO

2

value = actual CO

2

value x 0.952

Note

If the ETCO

2

value is displayed as + + +, verify calibration against a known

reference gas. If the sensor calibration is not accurate, return it to Welch Allyn for
service.