Printing

Welch Allyn Propaq Encore Vital Signs Monitor

OxyCRG

The OxyCRG is a graphical printout of two minutes of continuous HR/PR and SpO

2

numerics, and a condensed respiratory waveform. If any of the parameters have been
completely inactive for the two minutes prior to the initiation of the print, the associated
band will be empty.

OxyCRG on alarm

When an alarm condition is detected, a print will be initiated if OXYCRG ON ALARM is
turned on. If an SpO

2

or HR/PR alarm condition is detected, an OxyCRG will be queued to

print 60 seconds after the alarm is detected. If an Apnea or RR/BR alarm condition is
detected, an OxyCRG will be queued to print 75 seconds after the alarm is detected.

The parameters which have alarmed in the two-minute period are indicated by the
highlight of the corresponding labels.

A 3-second mark

Printout continues

Alarm occurred here

Alarming vital sign numerics are highlighted