5 ambient conditions, 6 electrical connection, 5 ambient conditions 4.6 electrical connection – FRITSCH ANALYSETTE 18 User Manual

Page 19: Installation

Advertising
background image

3.

Select an appropriate position for the wall bracket of the separate

control device provided, and secure the bracket in place.

4.

Connect the controller provided for the sieve shaker to the device

drive.

4.5 Ambient conditions

WARNING!
Mains voltage!
– The device may only be operated indoors.
– The surrounding air may not carry any electrically con-

ductive dust.

– Maximum relative humidity 80% for temperatures up

to 31°C, linearly decreasing down to 50% relative

humidity at 40°C.

n

The room temperature has to stay between 5 - 40°C.

n

Altitudes up to 2000 m

n

Degree of pollution 2 according to IEC 664.

4.6 Electrical connection

DANGER!
Provide short-circuit protection!
Risk of damage due to short-circuits.
– Make sure that the socket is connected to a mains line

protected with a residual current circuit breaker.

DANGER!
Mains voltage!
Changes to the connection line may only be made by a

qualified technician.

Installation

- 19 -

Advertising