Of 300 and 400 mm, Using the device – FRITSCH ANALYSETTE 18 User Manual

Page 23

Advertising
background image

n

DIN 66 165, part 1 basics - 1987 version

n

DIN 66 165, part 2 procedure - 1987 version

The standard sheets can be purchased from Beuth-Verlag, Berlin.

5.3 The maximum particle size permitted for test sieves

To avoid damaging the sieve base, the particle dimensions in a batch

should not exceed 10 w

0.7

(w is mesh width in mm).

Example

Nominal mesh width, w

Maximum particle size

mm

approx. mm

0.045

1

0.25

4

1.00

10

4.00

25

5.4 Recommended bulk volume for test sieves with diameters of 300 and 400 mm

Bulk volume for test sieves with a diameter

of 300 mm

Nominal mesh width

w

Feeding volume

Maximum sieve res-

idue

µm

approx. cm

3

approx. cm

3

25

33

17

32

39

20

45

45

23

63

53

26

90

65

32

125

75

38

180

90

45

250

105

53

355

120

60