3.4.4 CONFIGURATION: COUNT/PULSE.



NOTE

: Used with pulse input only. Each pulse in is

multiplied by the count/pulse factor.
1. Begin with the ENTER light OFF.
2. Turn the thumbwheel switch to position 7.
3. Press the ENTER switch. The ENTER light begins flashing.
4. Turn the thumbwheel switch to position 4.
5. Press the ENTER switch. The ENTER light turns on.
6. Turn the thumbwheel switch to position:
1 for 1
2 for 2
5 for 5
7. Press the ENTER switch. The ENTER light turns off.


------------------ COUNT/PULSE -----------------



FLASH ON OFF (Enter Light)
Set 1x 7 4 1 (Thumbwheel Switch)
Set 2x 7 4 2 (Thumbwheel Switch)
Set 5x 7 4 5 (Thumbwheel Switch

)








REMEMBER:

Return the thumbwheel switch to the correct mode position after configuration