2 serial connection of lan/wlan adapter, Installation and connection – VACUUBRAND VACUU·CONTROL User Manual

Page 25

Advertising
background image

Installation and connection

25

BA-N°: 999279/17.10.2014

4. Press the holder firmly on the pumping unit.

5. Insert the WLAN adapter into the attached holder.

4.2 Serial connection of LAN/WLAN adapter

CVC 3000 – VARIO pumping unit PC 3001 VARIO/VARIO

PRO

1. Switch off your pumping system,

ON/OFF switch in position

0.

2. Pull the CVC 3000 controller carefully out of the housing.

Be careful not to disconnect existing connections!

3. Feed the sub-D cable underneath the housing and between

the pump feet to the CVC 3000.

4. Connect the sub-D plug of the adapter to the sub-D

1

connec-

tion at the rear of the CVC 3000.

5. Fix the sub-D plug with its screws.

6. Insert the CVC 3000 controller into the housing.

5

CVC 3000 prepared.

CVC 3000 – as bench‑top device or for lab stand

Also valid for CVC 3000 mounted directly atop pumping units.

1. Switch off your CVC 3000.

2. Separate the CVC 3000 from mains (unplug mains plug) or 

unplug the connection cable to the VARIO pump.

3. Connect the sub-D plug of the adapter to the sub-D connec-

tion at the rear of the CVC 3000.

4. Fix the sub-D plug with its screws.

5. Reinstall the disconnected cable connections.

5

CVC 3000 prepared.

1 -> Sub-D = serial connection