3 reset to factory default settings (reset), Resolving problems, Caution – VACUUBRAND VACUU·CONTROL User Manual
Page 72

Resolving problems
72
BA-N°: 999279/17.10.2014
8.3 Reset to factory default settings (Reset)
CAUTION
Damage due to interrupted connection to the
vacuum system.
If
VACUU•CONTROL
®
adapter is reset to factory
default settings, contact to the vacuum system will be
lost. The remote control will stop functioning.
>
Stop any vacuum process prior to resetting to factory
settings.
>
Only authorized personnel may carry out a reset.
The remote control can be reset to factory settings directly at the
LAN or WLAN adapter. The latest
VACUU•CONTROL
®
Firmware
and GUI updates are preserved. Data like measured values in the
data logger or passwords etc. are deleted.
For resetting, the LAN/WLAN adapter has to be connected to the
controller CVC 3000.
Reset – WLAN adapter
1. Switch off pumping unit/CVC 3000 controller.
2. Use a straight, stable wire
1
, Ø = 0.8 ‑1 mm, l = 25 mm.
3. Insert the wire into the small opening at the LAN/WLAN
adapter.
4. Press the reset push-button and keep pressed for about
10 seconds while switching on the pumping unit/CVC 3000.
5
Perceptible movement of push-button,
5
VACUU•CONTROL
®
reset to factory settings.
Reset – LAN adapter
1. Use a straight, stable wire, Ø = 0.8 ‑1 mm, l = 25 mm.
2. Insert the wire into the small opening at the LAN/WLAN
adapter.
3. Press the reset push-button and keep pressed until both the
yellow and the green LED at the Ethernet connection switch
off simultaneously.
5
Perceptible movement of push-button,
5
VACUU•CONTROL
®
reset to factory settings.
1 -> Ø = diameter
LAN adapter
LED’s
WLAN adapter