4 parameter settings, Operation – VACUUBRAND VACUU·CONTROL User Manual

Page 48

Advertising
background image

Operation

48

BA-N°: 999279/17.10.2014

Meaning

Menu

control parameter

Pump down

6.4.4 Parameter settings

This button corresponds to the selection knob of the CVC 3000.

“Parameter” is a menu to fine‑tune the selected operation mode. 

Parameters can also be adjusted during operation.

Open menu “Parameter”

>

Click on the button with the gear-wheel icon.

5

Button marked,

5

“Parameter” menu is displayed.

The display

3

 of the “Parameter” menu depends upon the mode 

selected and the remote-control device.

Parameter – Pump down

Parameter

Meaning

Speed (%)

Only VARIO: motor speed parameter for pump 

down.

Adjustment range: 1 ‑ 101 (101 = HI mode)*

Minimum

(mbar)

Setting vacuum value which has to be reached.

When this value is reached, the vacuum controller

switches off or the in-line valve closes.

Adjustment range: 0; 1 ‑ 1060 (0 = OFF)

Duration (min)

Default value for process runtime from “Start“ on.

Adjustment range: 0 ‑ 1440 (0 = OFF)

Delay (min)

Delay time for an optional coolant valve.

* HI mode: optimum speed for the respective pressure, recommended setting.

3 -> Screenshots of “Parameter“ are examples with VARIO pump with pressure

unit in mbar.

Advertising