6 venting the system (vent), Operation, Danger – VACUUBRAND VACUU·CONTROL User Manual

Page 53

Advertising
background image

Operation

53

BA-N°: 999279/17.10.2014

6.4.6 Venting the system (Vent)

DANGER

Danger of explosion when venting with air by

forming explosive mixtures.

Depending on the process venting can form explosive

mixtures.

>

Never vent processes with air which can form an

explosive mixture.

>

If necessary vent with inert gas (max. 1.2 bar abs.).

IMPORTANT! Certain process may cause pressure to build

up.

The “Vent” button is used to vent the system. A short click on this 

button will momentarily vent the system as the process contin-

ues. Holding the vent button for longer than 2 seconds will cause

the system to be vented to atmospheric pressure and the pump

will stop running.

Momentarily venting

>

Click on the “Vent” button less than 2 seconds.

5

Button momentarily marked,

5

venting impulse, venting valve opens momentarily

short-term pressure increase.

Continuous venting

>

Click on the “Vent” button longer than 2 seconds.

5

Button is marked as long as pressed,

5

symbol for venting valve is blinking,

5

venting valve open  continuous pressure increase until

atmospheric pressure.

>

When atmospheric pressure is reached, the venting valve

closes automatically.

Advertising