SECTION 15 9600 SPECIFICATIONS

15.1 SENSOR SPECIFICATIONS


USING 2 MM PATHLENGTH DETECTOR CELL

Range: 0 to 10 mg/L Nitrate-N

Accuracy: 0-6 mg/L -- +/- 5 % of reading or 0.2 mg/L, whichever is greater.

Accuracy: 6-10 mg/L -- +/- 10 % of reading using a 6 mg/L Nitrate-N standard. Accuracy in the
0-6 mg/L range may be slightly compromised by the use of the 6 mg/L standard.

Detection Limit:
0.025 mg/L Nitrate-N



USING 10 MM PATHLENGTH DETECTOR CELL

Range: 0 to 2 mg/L Nitrate-N

Accuracy:

+/- 5 % of reading or 0.02 mg/L, whichever is greater


Detection

Limit:

0.005 mg/L


15.2 SYSTEM SPECIFICATIONS



Operating Temperature Range: 1 to 45 C

Operating Depth : < 200 feet (61 m) of water

Reagent Life: > 30 days at a 1 hour sample interval

Standard Battery Configuration: YSI 9611 Battery Pack with 14 Alkaline D Cells.

Battery Life: Approx. 70 days at a 1 hour sample interval at 20 C; approximately 30 days at a 1 hour
sample interval at 5 C.

Dimensions: Cylindrical with a diameter of 13 inches and a height of 28 inches.

Weight With All Reagent Bags Full and Alkaline Battery Pack Installed: Approximately 40 lbs.
(18.2 kg).

Communication Protocol:

RS-232 for communication to PC


SDI-12 for communication to Data Collection Platforms


YSI Environmental

86

Model 9600 Nitrate Monitor