4

• Introduction

1.2

Applications

Industrial and laboratory ON/OFF control

Energy management

Annunciation

16 TTL/DTL compatible digital input channels

Security controller

Product test

Period and pulse width measurement

Event and frequency counting

Waveform and pulse generation

BCD interface driver

1.3

Specifications

Analog Input (A/D)

Converter: ADS774 or equivalent, successive approximation type

Resolution: 12-bit

Number of channels:

ACL-8112DG/HG: 16 single-ended or 8 differential

ACL-8112PG: 16 single-ended

Input Range: (Programmable)

ACL-8112HG:

Bipolar :

±

10V,

±

5V,

±

1V,

±

500 mV,

±

100mV,

±

50mV,

±

10mV,

±

5mV

Unipolar: 0~10V, 0~1V, 0~0.1V, 0~0.01V

ACL-8112DG:

Bipolar :

±

10V,

±

5V,

±

2.5V,

±

1.25V,

±

0.625

Unipolar: 0~10V, 0~5V, 0~2.5V, 0~1.25V