Calibration & utilities – ADLINK ACL-8112 Series User Manual

Page 57

Advertising
background image

Calibration & Utilities

• 49

6

Calibration & Utilities

With data acquisition processes, knowing how to calibrate your
measurement devices to maintain its accuracy is very important. Users
can calibrate the analog input and analog output channels under user’s
operating environment to optimize the accuracy of the equipment. This
chapter will guide you through how to calibrate the ACL-8112.

6.1

What do you need

Before calibrating your ACL-8112 card, you should prepare some
equipment for the calibration:

Calibration program: The calibration program 8112util.exe. Once the
program is executed, it will guide you through the calibration
process. This program is included in the delivered package.

A 5 1/2 digit multimeter (6 1/2 is recommended)

A voltage calibrator or a very stable and noise free DC voltage
generator.

Advertising