11 span calibration – ADLINK ND-6013 User Manual
Page 56
                                    Advertising
                                    
                                        
                                    
                                
                                 
48
• Command Set
3.11 Span Calibration
@Description
To correct the gain errors of AD converter by using the span calibration.
@Syntax
$(Addr)0<CR>
$
Command leading code (1 character)
(Addr)
Address ID (2 character)
0
Span calibration (1 character)
@Response
!(Addr)<CR> 
 or 
?(Addr)<CR
> 
 
!
Command is valid.
?
Command is invalid.
(Addr)
Address ID.
@Example
User command:
$060<CR>
Response:
!06<CR>
 
To perform the span calibration for analog input module, address ID is 
06H. 
 
Note:
To perform the calibration, a proper input signal should be 
connected to the analog input module. Different input range 
have different input voltage, detail refer chapter 5 “Calibration”. 
(6011/D, 6012/D, 6013 C4.6
6014D, 6017, 6018)