Chapter 5: Using Your Enclosure
Product Service 1-800-522-7658
23
Figures 5-8
Only load materials required for the procedure. Do not overload the
enclosure with obstructions that might disrupt airflow.
Large objects should not be placed close together and should be elevated at
least ¼" above the work surface to permit airflow to sweep under the
equipment.
Secure both inner bags of the source container to the retaining ring with
retaining bands as shown in Figure 5-8.
After loading and attaching the bags, wait 15 seconds to purge airborne
contaminants from the work area. Do not remove your hands from the
enclosure.
For small containers, install the access opening cover in place of the
retaining ring.
Work Techniques
Keep all materials behind the lower air foil, and perform all contaminated
operations as far to the rear of the work area as possible.
Segregate all clean and contaminated materials in the work area.
Avoid using techniques or procedures that disrupt the airflow pattern in the
enclosure.
Work Procedure
Carefully transfer powder from the source container with scoop properly
sized for the receiving container.
When the desired amount is reached, close the receiving containers and
place to the side. Repeat process for each. The user may either wipe
container and lid after each transfer or when all transfers are complete.
Inner Bag
Retaining
Band