Chapter 5: Using Your Enclosure

Product Service 1-800-522-7658

23

Figures 5-8

Only load materials required for the procedure. Do not overload the

enclosure with obstructions that might disrupt airflow.

Large objects should not be placed close together and should be elevated at

least ¼" above the work surface to permit airflow to sweep under the
equipment.

Secure both inner bags of the source container to the retaining ring with

retaining bands as shown in Figure 5-8.

After loading and attaching the bags, wait 15 seconds to purge airborne

contaminants from the work area. Do not remove your hands from the
enclosure.

For small containers, install the access opening cover in place of the

retaining ring.

Work Techniques

Keep all materials behind the lower air foil, and perform all contaminated

operations as far to the rear of the work area as possible.

Segregate all clean and contaminated materials in the work area.

Avoid using techniques or procedures that disrupt the airflow pattern in the

enclosure.

Work Procedure

Carefully transfer powder from the source container with scoop properly

sized for the receiving container.

When the desired amount is reached, close the receiving containers and

place to the side. Repeat process for each. The user may either wipe
container and lid after each transfer or when all transfers are complete.

Inner Bag

Retaining
Band